Volume Range and Accuracy Specifications
1. Volume Range: 0.5 - 10 µL
Increment: 0.1 µL
Test Volume (µL): 10
Accuracy Error: 1.50% (0.15 µL)
Precision Error: 1.50% (0.15 µL)
Test Volume (µL): 5
Accuracy Error: 2.50% (0.125 µL)
Precision Error: 2.50% (0.125 µL)
Test Volume (µL): 1
Accuracy Error: 4.00% (0.04 µL)
Precision Error: 4.00% (0.04 µL)
2. Volume Range: 5 - 50 µL
Increment: 0.5 µL
Test Volume (µL): 50
Accuracy Error: 1.00% (0.5 µL)
Precision Error: 0.50% (0.25 µL)
Test Volume (µL): 25
Accuracy Error: 1.50% (0.375 µL)
Precision Error: 1.00% (0.25 µL)
Test Volume (µL): 5
Accuracy Error: 3.00% (0.15 µL)
Precision Error: 2.00% (0.1 µL)
3. Volume Range: 50 - 300 µL
Increment: 5 µL
Test Volume (µL): 300
Accuracy Error: 0.70% (2.1 µL)
Precision Error: 0.25% (0.75 µL)
Test Volume (µL): 150
Accuracy Error: 1.00% (1.5 µL)
Precision Error: 0.50% (0.75 µL)
Test Volume (µL): 50
Accuracy Error: 1.50% (0.75 µL)
Precision Error: 0.80% (0.4 µL)